Message: | A:Brief description:
DSHD-265D-1 Kinematic Viscometer is applied to determine the kinematic viscosity of liquid petroleum products (Newton liquid) under a fixed temperature environment as per GB/T 265 Standard Test Method for Kinematic Viscosity and Calculation Method for Dynamic Viscosity of Petroleum Products and ASTM D445.
B:Technical advantages
1.The instrument is a kind of specially-made testing machine which the temperature control accuracy can reach ±0.01℃ and having digital display.
2.It adopts hard glass vessel and heat preservation cover(double-layer vessel).Good heat prevention and easy to observe the sample.
3.It adopts desktop and all-in-one machine design,convenient to use.
4.It adopts electric stirrer to ensure the uniform of bath temperature.
C:Technical specifications and parameters
1.Power supply: AC 220 ± 10 % 50 Hz.
2.Heating power: Auxiliary heater: 1000 W; Temperature Control heater: 600 W.
3.Power for stirrer: 6W, rotating speed:1200RPM
4.Temperature range: ambient~100 ℃.
5.Temperature control accuracy: ±0.01℃.
6.Timer: 0.0s~999.9s
7.Constant temperature bath: 20L, double layer.
8.Ambient temperature: -10℃~+35℃
9.Relative humidity: ≤85%
10.Total power consumption: not higher than 1800W.
11.Capillary viscometer: 7 pieces in total. The inner diameters are 0.6, 0.8,1.0, 1.2, 1.5, 2.0 and 2.5mm
12.Overall dimension: 530mm*400mm*670mm(water bath is included) |