IgG IgM Rapid Test

Create: 07/18/2020
Expired Date:07/19/2022
Category: Electronic
Message:1. Purpose The purpose of this proposal is to provide an validation of the analytical sensitivity. The production of at least three consecutive 2019-nCOV IgG/IgM Rapid Test Device products shall be controlled. 3. Material Positive Control:2019-nCOV-IgM; Positive Control:2019-nCOV-IgG; Validation lot 1:NO1G01T; Validation lot 2:NO1G02T; Validation lot 3:NO1G03T. 4. Method Tests the low positive control of 2019-nCOV-IgM,2019-nCOV-IgG and negative sample.Each specimen tests in 11 tests.Read the positive result at 10 min.Do not interpret the result after 15 minutes. 5. Interpretation of results Positive result:≥G3 Negative result:<G3 6. QC Acceptance Criteria Serum/Plasma specimens:C-line≤3min. Whole blood specimens:C-line≤5min. C-line≥G8 in 10min. The 2019-nCOV IgG/IgM Rapid Test Device can show good performance using the positive and negative samples. url:http://www.codewel.com